pH Transmitter Comparison Chart
4-Wire Transmitters 2-Wire Transmitters
Model Model 10 Model 12 Model 14 Model 16 Model TB82
SENSOR INPUT
Number of pH Inputs 1 1 1 1 1
Separate Temperature Input YES YES YES YES YES
TRANSMITTER OUTPUT
Number of Outputs 1 2 2 1 1
Isolated YES YES YES YES YES
0/4-20 mA Output Configuration YES YES YES YES YES
INSTRUMENT MEASUREMENT SPEC
Measurement Range 0 to 14 pH -2 to 16 pH 0 to14 pH -2 to 16 pH -2 to 16 pH
Temp Range (measuring) -15 to 100° C -30 to 140° C -20 to 200° C -30 to 140° C -20 to 300° C
Storage Temp (meter) -10 to 65° C -30 to 70° C -20 to 70° C -30 to 70° C -40 to 70° C
Output Repeatability ± 0.01PH ± 0.01 pH ± 0.02 pH ± 0.01 pH ± 0.01 pH
Stability ± 0.01 pH / month ± 0.01 pH ± 0.01 pH ± 0.01 pH ± 0.01 pH
Removable Terminal Strips NO NO Yes power, No sensor NO YES
Solution Ground NO YES YES YES YES
HART Protocol NO YES NO YES OPTIONAL
TEMPERATURE COMPENSATION
Automatic YES YES YES YES YES
Manual YES YES YES YES YES
Compatible RTD Pt 100 Pt 100, Pt 1000, 3K Pt100, Pt1000 Pt 100, Pt 1000, 3K Pt100, 3K Balco
DISPLAY Type LED LCD LCD LCD LCD
Dual (pH & T) NO YES YES YES YES
Resolution ± 0.01 pH ± 0.01 pH ± 0.01 pH ± 0.01 pH ± 0.01 pH
CONFIGURATION AND KEYPAD ENTRY
Scrolling/branching Menus YES YES YES YES YES
Yes/No Dialog Menus NO YES YES YES YES
Alphanumeric Display NO YES YES YES YES
Mnemonic Prompting YES NO NO NO NO
Complete Word Prompting NO YES YES YES YES
FRONT PANEL & SECURITY
Membrane Keypad YES Buttons YES Buttons YES
User Selectable Security YES YES YES YES YES
Without Security Codes Read only Read Only Read only Read Only Read Only
Secured Access Level 1 All access, calibrate, display, service Calibrate pH, set display, select manual or auto temp comp. Calibration-pH & temp. Commission-Cal pH, set display, set manual or auto TC Calibrate, set holds, Configure
Secured Access Level 2 N/A Change ranges, set points, holds & alarms Program measurement alarms, output Change ranges, set points, holds & alarms Calibrate, set holds
Secured Access Level 3 N/A Activate or configure wash timer Change display, independences and access codes Activate or configure wash timer N/A
Independent Access Levels or Tiered N/A Either Independent Either N/A
Override Instructions in Manual YES NO YES NO Reset only
ENCLOSURE CLASSIFICATION
NEMA 4X and IP65 YES YES YES YES YES
MOUNTING
Included Hardware Panel Panel Panel, Pipe and Wall Panel Panel
Options Available Pipe Pipe and Wall _______ Pipe and Wall Pipe and Wall
CALIBRATION
Semi-automatic or Buffer Recognition YES YES YES YES YES
Manual Mode for Grab Sampling YES YES YES YES YES
Error Messages Codes Codes Text Codes Codes and Text

Back to top

Home | FAQs | Documents | Tech Support | Contact | Site Map | Search

Broadley-James Corporation Toll-Free 800.288.2833
Copyright © 2001 Broadley-James Corporation. All Rights Reserved.