Model SA 877PVC — DynaProbe® Sensor Housing for Valve Retraction Applications
Specifications
Ordering Information

The Model SA877PVC is a sensor housing assembly designed for use with the replaceable Model ST877 and ST977 DynaProbe sensors. This housing is suited for the most severe chemical industry applications where metals cannot withstand the corrosive nature of the process and is intended for valve retraction service. The Coupler is constructed from chemically resistant Kynar (PVDF). PVC is the standard material of construction for all other parts of this assembly.

Insertion length is 7 inches. Viton O-ring seals are incorporated throughout this housing to prevent damage to the sensor and to retard process seepage during insertion and retraction procedures. The sheath is designed with blowout protection, should the compression fitting be loosened at the process operating pressure. The SA877PVC sensor housing is designed with replaceable components providing simple and economic servicing.

SPECIFICATIONS

pH Range

0-14 pH (w/DynaProbe)

ORP Range

± 5000 mV (w/DynaProbe)

Temp Range

0º to 55º C

Max Pressure/Temp

50 psig at 50º C

Wetted Materials

PVC, Kynar (PVDF), and Viton O-Rings

ST877 and ST977 DynaProbes must be ordered separately

Back to top

Ordering Information
Call 1-800-288-2833 — For Price and Delivery

Specify Model SA877 Part Number here:

1
2
3
4
5
6
7
SA877—
 P
L  
0
 7
 —
J
 
 
Category Selection:
1 — 4 Sheath Material and Insertion Length:
PL07 PVC inches – standard material and length
5 — 7 Junction Box:
[J0P] Optional PVC junction box subassembly
[J99] Junction box not required – standard design

DynaProbe® pH and ORP Sensors
Must be Ordered Separately

Select the DynaProbe® sensor for pH or ORP (Redox) measurement which is compatible with the host pH/ORP instrument. A wide selection of DynaProbe® sensors both with and without automatic temperature compensation are available for most makes and models of pH/ORP instruments. For factory assistance, call our Tech Services at 1-800-288-2833.


Preamplifiers are also Available!
Ordered Separately

Select the optional preamplifier module to provide the correct interface signal between the DynaProbe® sensor in the Model SA877 housing and the host pH instrument. A wide selection of preamplifier modules both with and without junction boxes are available for most makes and models of pH instruments.