The ST856 is a rugged, sealed sensor assembly
designed for in-line or submersion applications. The patented solid state reference cell features the unique IonTrap™ design for extended service life in the most severe applications. The body, with an integral 3/4" MNPT insertion thread, is molded from chemically resistant Kynar and the reference junction is porous Teflon. Built-in temperature compensators are available. Features include built-in sensor guard that protects the sensor tip from impact. Sensor also available in ORP (Redox) version.

SPECIFICATIONS

pH Range

pH Range

ORP Range

± 5000 mV

Temp Range

0 -100º C

Max Pressure/Temp

100 psig at 50º C

Reference

Ag-AgCl

Wetted Materials

Kynar, Teflon, Glass

Standard cable length

7 Feet




Back to top

Ordering Information
Call 1-800-288-2833 — For Price and Delivery

Before you call to place your order, please print out this page and fill in this PART# BOX by selecting from options below. Boxes 1-9 must be filled in.

Part #
1
2
3
4
-
5
6
7
8
9
ST856-
-
-
-
-
-
-
-
-
-
OPTIONS Check options 1-9 below to complete PART# BOX above.
1 — Liquid Junction
[W] Coaxial Wood
[R] Recessed Coaxial Teflon
2-3 — Measurement Electrode
[33] pH, HT-3 Rugged Dome Bulb
[43] pH, HT-4 High pH bulb (for pH 13 and above)
[PB] ORP, Platinum Band
4 — Temperature Compensator: (for direct input into following manufacturer's preamp/instrument)
[0] T.C. not required
[1] 3K Ohms for most older Models of TBI/Bailey, Uniloc, and other US manufacturers
[9] 301 Ohms for older Great Lakes Models: 60, 70, 90, 570, and 690
[D] Pt 100 Ohms for Beckman Model SoluComp, TBI/Bailey (ABB) TB82 & TB84, Uniloc 1054, Foxboro 870 & 874, Yokogawa, and most European Models
[H] Pt 1OOO Ohms for Yokogawa, Conducta, Polymetron, LTH, & many European models
5 — Cable Type:
[A] Low Noise Coax - for more than 7 ft.
[E] Jacketed Coax w/ TC leads - for more than 7 ft.
[T] Teflon Coax ( > 80º service ) for more than 7 ft.
[U] Jacketed Teflon Coax w/TC leads ( > 800 service ) for more than 7 ft.
6-7 — Cable Length:
[07] 7 ft. is standard length
[XX] Specify length (in feet) above 7 ft. standard length
8-9 — Required Sensor Connectors:
[BC] BNC only (for use with ORP electrodes or with sensors w/o TC or solution ground)
[FB] BNC for pH and ferrules on all other leads (for use with Terminal Block or Terminal Strip)
[FF] All Ferrule connectors ( for use with Terminal block or Terminal Strip)

Back to top

Home | FAQs | Documents | Tech Support | Contact | Site Map | Search

Broadley-James Corporation Toll-Free 800.288.2833
Copyright © 2004 Broadley-James Corporation. All Rights Reserved. Trademarks.