SPECIFICATIONS
|
| pH Bulb: |
HT-3 Formula, Low Sodium Ion Error |
| Reference: |
Silver/Silver Chloride (Ag/AgCl) |
| Reference Junction: |
Precision made junction of specially formulated
low porosity ceramic |
| Electrolyte: |
3.8 Molar Potassium Chloride (KCl) Gel |
| Isopotential Point: |
pH 7 |
| Output Signal per pH Unit @ 25ºC: |
Approximately 59 millivolts |
| Assymetry Potential: |
0mV ±20mV |
Temperature Range
SP200 Series: |
Operation: -5 to 110ºC |
| Stability: |
Drift of less than 0.002 pH over 24 hours
at constant temperature & pressure |
| Response Time at 98% of readout in: |
25ºC less than15 seconds |
| Response Time at 98% of readout in: |
37°C less than 10 seconds |
| Max Pressure SP200 Series: |
10 bar (atm), 150 psig |
| Connector: |
Standard Pg13.5 threaded cap with -S8 cable
attachment. O-ring sealed |
Wetted Materials
SP200 Series: |
Glass, Ceramic |
Under laboratory Conditions
For High Pressure and/or High Temperature applications, request
literature on Broadley-James' DynaProbe® sensor line. |